24
NewTom GO 2D/3D
EN
2D/3D OPT E.L. CALIBRATION MANUAL
•
DIAMETER 8
Select the 8x10 field, enter the -17 and -18 values in
the “Left” and “Right” fields and press “move”. As soon
as the machine is ready, perform a new acquisition (8
mA and 20 ms).
Evaluate the image quality and, if necessary, change
collimator position until obtaining an image like the fol-
lowing one (...)
Once finished, press “Save” to save the new setting.
Then, select and save also for the following view
fields:
•
7x7
•
8x6
•
8x10 HiRes
•
8x7 HiRes
•
8x6 HiRes
•
DIAMETER 6
Select the 6x7 field, enter the -31 and -28 values in
the “Left” and “Right” fields and press “move”. As soon
as the machine is ready, perform a new acquisition (8
mA and 20 ms).
Evaluate the image quality and, if necessary, chan-
ge collimator position until obtaining an image like the
following one