22
NewTom GO 2D/3D
10
EN
2D/3D OPT E.L. CALIBRATION MANUAL
The aim of the procedure is to make the area hit by the
rays (in grey) match with the green box delimiting the
view field. In addition, the shadow area at the edges
must be within the red lines (see next image).
Before continuing, make sure that the scan area is
free from any object.
•
DIAMETER 10
•
• Select the 10x10 field, enter the values -3 and -7
inside the “Left” and “Right” fields, then press “Move”.
Enter the 8 and 10 values in the mA and time fields,
then press “Acquisition” to perform a ray acquisition.
If the resulting image is like the following one