20
NewTom GO 2D/3D
EN
2D/3D OPT E.L. CALIBRATION MANUAL
4. 3D CALIBRATION
4.1 RAY BEAM CALIBRATION
From NNT main menu, press Ctrl+Alt+Shift+End keys
together to set software in service mode (“Service Le-
vel” will be displayed at the bottom left, under NNT)
From the “Tools” menu, select “Scanner tests”,
press “Reset” to set device in 3D mode.
Then, from the “Tools” menu, select “Beam limiter
test”,