13
NewTom GO 2D/3D
F
G
H
I
EN
2D/3D OPT E.L. CALIBRATION MANUAL
• Vertical alignment
Slightly loosen the following screws (F, G, H, I).
Correct the vertical alignment manually; tighten the
screws, select “Image not centred” and press “ok”.
Repeat the procedure until a vertically-centred image
is obtained.
When the image is perfectly centred, select “Image
centred” and press OK; the software will automatically
check the proper alignment even for the Bitewing and
Children panoramic positions.
At the end of the procedure, press Ok to confirm and
save the data.