33
NewTom GO 2D/3D
EN
2D/3D OPT E.L. CALIBRATION MANUAL
performs a complete acquisition of the QA phantom;
once finished, when system prompts you to continue
the analysis, press “Yes”.
The calculations are made automatically; once fini-
shed, a report like the one shown in the figure will be
displayed:
If no value has asterisks, close the window; should
this not be the case, check QA phantom positioning
and repeat the procedure.
4.5 CALIBRATION DATA BACKUP
From NNT main page, select “Setting Backup”
Select the folder for the Backup and press OK