April 2018
58
FPM-8220
TERM?
Common
Device Dependent
Front Panel
Action
Requests the currently defined message terminator.
Response
0 through 6, interpreted as follows:
0 =
<CR><NL><^END>
1 =
<CR><NL>
2 =
<CR><^END>
3 =
<CR>
4 =
<NL><^END>
IEEE-488.2 Standard and FPM-8220 default
5 =
<NL>
6 =
<^END>
Notes
All settings other than TERM 4 are out of compliance with IEEE-488.2 specifications. They are
provided for flexible compatibility with various GPIB drivers.
See TERM (above) for more information.
Examples
Term? - Response 4 means the message terminator is the FPM-8220 default: <NL><^END>
TIME?
Common
Device Dependent
Front Panel
Action
Requests the time since the last time the FPM-8220 was powered ON.
Response
ASCII character data in the form hours:minutes:seconds
Notes
Maximum time (clock "turns over") is1193 hours (~50 days).
The TIME clock is independent of the TIMER clock. See TIMER? below.
Examples
Time? - Response 0:32:01.76 means 32 minutes and 1.76 seconds have elapsed since power-ON.
TIMER?
Common
Device Dependent
Front Panel
Actions
Requests the time since the last TIMER? query. Resets the TIMER clock to zero.
Response
ASCII character data in the form hours:minutes:seconds
Notes
Maximum time (clock "turns over") is 1193 hours (~50 days).
The TIMER clock is independent of the TIME clock. See TIME? above.
The TIMER clock is set to zero at power-ON, so the first response will be the same as a TIME?
query.
Example
s
Timer? - Response 0:00:12.07 means 12.07 seconds have elapsed since the last TIMER? query.
Summary of Contents for FPM-8220
Page 1: ...User s Guide Fiber Optic Power Meter FPM 8220 70044001 April 2018...
Page 2: ......
Page 6: ...April 2018 iv FPM 8220 This page was intentially left blank...
Page 12: ...April 2018 x FPM 8220 This page was intentionally left blank...
Page 16: ...April 2018 4 FPM 8220 Specifications...
Page 34: ...April 2018 22 FPM 8220...
Page 78: ...April 2018 66 FPM 8220...