April 2018
50
FPM-8220
ENABle:COND <nrf value>
Common
Device Dependent
Front Panel
Action
Enables bits in the device condition status enable register.
Paramters
0 to 65535
Notes
Enabled/disabled
conditions
can be read by ENABle:COND?.
Changing condition status may be monitored by COND?.
Bit 3 of the status byte register is set if any device enabled conditions are true.
See Chapter 3 for more information about register structure.
Enable registers normally retain their values through power OFF/ON, unless the power-on status
clear flag is set true. See *PSC.
Examples
ENAB:COND 12 - Enables the condition status register so that over-range and under-range error
conditions are summarized in the status byte bit 3 (8 + 4 = 12).
Enable:cond #HC - Same as ENAB:COND 12, except using hexadecimal numbering. See
RADix.
ENABle:COND?
Common
Device Dependent
Front Panel
Action
Requests the value in the device condition status enable register.
Response
0 to 65535
Notes
Enabled/disabled conditions can be set by ENABle:COND?. Changing condition status may be
monitored by COND?. See Chapter 3 for more information about register structure.
Examples
ENAB:COND? - Response 4 means that over range will be reported in status byte bit 3.
Enable:COND? - Response #H4 is the same as 4, except using hexadecimal numbering. See
RADix.
Numeric
Value
Over-range
Under-range
4
8
= Unused bits
Measurement data ready
2048
Numeric
Value
Over-range
Under-range
4
8
= Unused bits
Measurement data ready
2048
Summary of Contents for FPM-8220
Page 1: ...User s Guide Fiber Optic Power Meter FPM 8220 70044001 April 2018...
Page 2: ......
Page 6: ...April 2018 iv FPM 8220 This page was intentially left blank...
Page 12: ...April 2018 x FPM 8220 This page was intentionally left blank...
Page 16: ...April 2018 4 FPM 8220 Specifications...
Page 34: ...April 2018 22 FPM 8220...
Page 78: ...April 2018 66 FPM 8220...