April 2018
42
FPM-8220
Command Reference
The following pages contain a reference for both common and device-dependent commands of the
FPM-8220 Fiber Optic Power Meter. This reference contains useful information for both local and
remote operation of the FPM-8220.
*CAL?
Common
Device Dependent
Front Panel
Action
Adjusts the internal analog to digital (A/D) converter to reference points, then reports results.
Response
Zero = OK
Non-zero = calibration error
Notes
The A/D chip in the FPM-8220 performs automatic calibration. Thus, this query always return 0.
Example
*CAL?
*CLS
Common
Device Dependent
Front Panel
Action
Clears status event registers: Standard Event Status, Device Event Status, and Error Queue.
Notes
Useful to clear registers before enabling service requests (SRQ).
Example
*CLS
*ESE <nrf value>
Common
Standard Event Status Enable
Device
Dependent
Front Panel
Action
Enables bits in the Standard Event Status Enable Register.
Parameters
The value must be between 0 and 255.
Notes
Bit 5 of the Status Byte Register (STB) is set if any enabled conditions are true.
Setting Bit 0 allows for generation of service requests from overlapped commands as previous
operations complete. This may be useful for ensuring that an operation is complete before starting a
measurement.
Examples
“*ESE 40” –action: Sets the Standard Event Status Enable Register to enable Bit 5 of the Status Byte
Register if a device-dependent error or a command error occurs (40 = 2
3
+ 2
5
).
Summary of Contents for FPM-8220
Page 1: ...User s Guide Fiber Optic Power Meter FPM 8220 70044001 April 2018...
Page 2: ......
Page 6: ...April 2018 iv FPM 8220 This page was intentially left blank...
Page 12: ...April 2018 x FPM 8220 This page was intentionally left blank...
Page 16: ...April 2018 4 FPM 8220 Specifications...
Page 34: ...April 2018 22 FPM 8220...
Page 78: ...April 2018 66 FPM 8220...