April 2018
43
FPM-8220
*ESE?
Common
Standard Event Status Enable Query
Device
Dependent
Front Panel
Action
Requests the value of the Standard Event Status Enable Register.
Parameters
None.
Notes
Bit 5 of the Status Byte Register is set if any enabled conditions are true.
Response is the sum of the enabled bits and must be a value between 0 and 255.
Examples
“*ESE?” –response: 68, meaning the User Request and Query Error bits have been enabled in the
Standard Event Status Enable Register (68 = 2
2
+ 2
6
).
*ESR?
Common
Standard Event Status Register Query
Device
Dependent
Front Panel
Action
Requests the value of the Standard Event Status Register.
Parameters
None.
Notes
Response is the sum of the enabled bits and must be a value between 0 and 255.
Allows for the determination of which type of error has occurred.
Examples
“*ESR?” –response: 32, meaning a command error has occurred.