FPM-8220
REF?
Common
Device
Dependent
Front Panel
Action
Requests the reference level value.
Response
GPIB-standard nrf value, in the units currently defined. See
MODE
.
Notes
Same function as pressing
dBm/W
and
REF
together on the front panel.
Examples
REF? - Response 0 means the reference level is 0 dBm.
Ref? - Response -20 means the reference level is -20 dBm (10
μ
W).
Ref? - Response 2.795E-006 means the reference level is 2.795
μ
W.
RESP?
Common
Device Dependent
Front Panel
Action
Requests the calibrated detector responsivity for the currently selected wavelength.
Response
GPIB standard <nrf value>, in units: mA / mW
Notes
The information should correspond with that on your latest ILX Lightwave calibration certificate for this
meter. If not, contact ILX Lightwave Customer Service to resolve the discrepancy.
This information can be useful for understanding the characteristics of your FPM-8220 meter. Notice
for example that detector current is much less at shorter wavelengths.
For more information, see the discussion of the calibration certificate at the end of Chapter Two.
Examples
RESP? - Response 1.193E-2 means the detector produces 0.01193 mA/mW at the currently set
wavelength.
TERM <nrf value>
Common
Device Dependent
Front Panel
Action
Defines the message terminator.
Paramters
0 =
<CR><NL><^END>
1 =
<CR><NL>
2 =
<CR><^END>
3 =
<CR>
4 =
<NL><^END>
FPM-8220 default and IEEE-488.2 Standard
5 =
<NL>
6 =
<^END>
Notes
All settings other than TERM 4
are technically out of compliance with IEEE-488.2 specifications.
However they are provided for flexible compatibility with various GPIB drivers.
TERM 3 will often cause "bus hang" problems, because IEEE-488.2 compliant instruments ignore the
<CR> as white space, causing them to wait indefinitely for message termination.
Meanings of the acronyms are historical:
CR = Carriage Return (e.g.: a mechanical typewriter)
NL = New Line (e.g.: line-feed on a "line printer")
^END = EOI code (End Or Identify)
Example
s
Term 4 - Define the message terminator as the IEEE-488.2 standard: <NL><^END>
Summary of Contents for FPM-8220
Page 1: ...User s Guide Fiber Optic Power Meter FPM 8220 70044001 April 2018...
Page 2: ......
Page 6: ...April 2018 iv FPM 8220 This page was intentially left blank...
Page 12: ...April 2018 x FPM 8220 This page was intentionally left blank...
Page 16: ...April 2018 4 FPM 8220 Specifications...
Page 34: ...April 2018 22 FPM 8220...
Page 78: ...April 2018 66 FPM 8220...