These specifications assume a
maximum load of 34 pounds (15.5 kg)
, including platforms,
clamps, glassware and contents.
Innova 43 & 43R Incubator Shaker
SHAKING
1
20-500 rpm with 1-inch (2.5 cm) stroke or 20-300 rpm with 2-inch (5.1 cm)
stroke
Speed
Control Accuracy
1 rpm
Displayed in 1 rpm increments
Indication
1-inch (2.5 cm) or 2-inch (5.1 cm)
Stroke/Orbit
TEMPERATURE
2
5ºC above ambient temperature to 80°C
Range (43)
2
20ºC below ambient temperature (minimum 4ºC) to 80°C
Range (43R)
Control Accuracy
0.1°C from 30-40 °C, 0.5°C over the remaining range.
Uniformity
0.25°C in flask across platform at 37ºC, and 0. 5°C (or less) at 60ºC
Indication
Displayed in 0.1°C increments
Heater
Long-life, low-watt density resistance-type heater with high temperature
thermostat
Visible and audible warning indication when speed deviates more than 5
rpm or temperature more than 1°C from setpoints, and when timer has
expired. Audible alarm can be muted.
ALARMS
240 x 128 backlit
LCD DISPLAY
Remote control, remote monitoring, remote data logging
RS-232
All setpoints and operating status are retained in non-volatile memory
SETPOINT RETENTION
Automatic restart after power is restored, indicated by flashing display
AUTOMATIC RESTART
Triple-eccentric counterbalanced drive with nine permanently lubricated ball
bearings and poly-V belt. Self-calibrating.
DRIVE
Solid-state brushless DC motor.
DRIVE MOTOR
Drive Interrupt shuts off power to Shaker when lid opens.
Acceleration/deceleration circuit prevents sudden starts and stops,
minimizing both splashing and mechanical damage.
SAFETY
100V, 50Hz
ELECTRICAL
REQUIREMENTS
100V, 60 Hz
43: 800 VA per shaker
120 Volts, 60 Hz
43R: 1500 VA per shaker
230 Volts, 50 Hz
UL 61010A-1
CAN/CSA-C22.2 No. 1010.1
ETL REGULATORY
STANDARDS
UL 61010A-2-010
CAN/CSA-C22.2 No. 1010.2.010
IEC 61010-1: 2001 2nd ed. EN 61010-2-081 (2002)
EN 61010-1: 2001 2nd ed. EN 61326 : 1997 + A1 : 1998
EN 61000-3-2 : 1995
CE REGULATORY
STANDARDS
EN 61000-3-3 : 1995
EN 61010-2-010 (1992) + A1: 1996
DIMENSIONS
3
53 inches (135 cm)
Width
32 inches (81 cm)
Depth (Front to Back)
40.6 inches (103 cm)
Height
1
see first NOTE on following page
2
depending on ambient factors, such as relative humidity
,
and options installed
3
see second NOTE on following page
New Brunswick Scientific Co., Inc.
User’s Guide
Summary of Contents for Innova 43