52
Figure 20: Calibrate Screen
CALIBRATE
°C Offset :
0.0
Calspeed :
250
CAL Th 16:22
This is a sample
of the
INDICATED RPM
VALUE.
7.8
Using Calspeed
The
Calspeed
function, which is set in the
CAL
screen (
see Figure 20 above
), is used to
calibrate the speed of the shaking mechanism. Calibrated at the factory, speed does not
need to be recalibrated until a major operating component (e.g., drive belt) is changed.
NOTE:
Prior to calibrating the speed, make sure that the platform is properly
secured to the subplatform, and any flasks present are secured.
When the shaker is running, the
CAL
screen shows the indicated RPM value. If you wish
to calibrate the speed, set the speed to a value that can be measured— a setpoint of 250
RPM works well. The use of a strobe is recommended for accuracy. If, after measuring
the actual speed, you wish to adjust the indicated value:
1. Click the Control Knob in.
2. Set the new value.
3. Click the Knob in again to save the setting.
4. Turn the shaker
OFF
, wait a few seconds, then turn the shaker back
ON
.
7.9
Power Interruption
In the event of a power failure, the Innova 43/43R Shakers are equipped with an
automatic restart
function. The shaker’s non-volatile memory retains all stored
information.
Innova 43 & 43R Incubator Shakers M1320-0050
User’s Guide
Summary of Contents for Innova 43
Page 2: ...2 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 10: ...10 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 79: ...79 11 1 1 CE Declaration of Conformity New Brunswick Scientific Co Inc User s Guide ...
Page 92: ...92 Filling the 33 Wavy Arrows Icon 21 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...