32
When the humidity sensor is present in the chamber,
maximum shaker temperature is
automatically limited to 60°C.
Specifications:
Humidity Operating Range:
0 to 100% RH
Temperature Operating Range:
4 to 60ºC
Accuracy:
See
Operating Range & Error Graph
, below
Hysteresis:
± 1.5% RH
Recovery Time after Saturation:
10 seconds
Long Term Stability:
0.5% RH per year
Figure 12b: RH Sensor Operating Range & Error Graph
Operating Range and Humidity Error
0
10
20
30
40
50
60
70
80
90
100
0
10
20
30
40
50
60
Temperature (C)
+/- 5% Error
+/- 3% Error
+/- 5% Error
INTERMITTENT
OPERATION
Figure 12c: RH Sensor Connection