83
12.4
Set Commands
Code
Meaning
Format
CL
Command Grow
Lamp
CL
<
Space
><
Arg1
><
CR
>
where
Arg1
= Set Grow Lamp state (0 = off, 1 = on)
CS
Command Speed
CS
<
Space
><
Arg1
><
CR
>
where
Arg1
= Agitation setpoint (####)
CT
Command
Temperature
CT
<
Space
><
Arg1
><
CR
>
where
Arg1
= Temperature setpoint (###.#)
CU
Command UV
Lamp
CU
<
Space
><
Arg1
><
CR
>
where
Arg1
= Set UV Lamp state (0 = off, 1 = on)
12.5
Profile Control Commands
Code
Meaning
Format
PC
Clear Profile
(clears profile
step)
PC
<
Space
><
Arg1
><
Space
><
Arg2
><
CR
>
where
Arg1
= Profile number (range 1 to 4) and
Arg2
=Step number (range 1 to 15)
NOTE:
If
Arg2
is not included, all 15 steps of the
selected profile will be cleared.
PM
Profile Mode
(returns status of
current profile
run/hold mode)
PM
<
CR
>
There are no parameters to this command.
Return data format:
<
Arg1
><
Space
><
Arg2
><
Space
><
Arg3
><
CR
>
where
Arg1
=Run/hold status (0 = end, 1 = run),
Arg2
=Profile number and
Arg3
=Step number
NOTE:
If a profile has not been selected to run, a 0
will be returned for Profile number and Step number.
...continued...
New Brunswick Scientific Co., Inc.
User’s Guide
Summary of Contents for Innova 43
Page 2: ...2 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 10: ...10 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...
Page 79: ...79 11 1 1 CE Declaration of Conformity New Brunswick Scientific Co Inc User s Guide ...
Page 92: ...92 Filling the 33 Wavy Arrows Icon 21 Innova 43 43R Incubator Shakers M1320 0050 User s Guide ...