60
Figure 26: Fuse Replacement
FUSES
(F1 – F4)
SERVICE
COMPARTMENT
9.5
Replacing Cabinet Lights
WARNING!
Before installing or removing the quartz bulbs:
(1) Make sure power is disconnected from the unit,
(2) Wear plastic gloves to protect the bulb from skin oils and to protect your
hands just in case the bulb should break,
(3) Before removal, make sure the bulb has fully cooled.
NOTE:
Cabinet lights are connected in series. When one burns out, both lights will
be off. It is therefore prudent to replace both bulbs at the same time.