3.
unit cannot be zeroed, refer to the
Troubleshooting
in
Appendix A.
4.
Turn on the ultrasound unit following the manufacturer’s instructions. Adjust it to
the desired level. Never exceed thirty watts.
5.
Read the actual output power in WATTS on the LCD of the UPM 2000.
6.
After all of the readings are completed, shut off the ultrasound unit.
ove the water from the transducer well when the measurements are complete.
ing the drain tube located in the bottom front, completely drain the well. Remove
g the tube end over a container, pinch the
h the
letely.
It is important
mplete. This will prevent bacterial growth
8.
ing it ineffective as a coupling medium.
Allow five minutes for the unit to warm up and stabilize before starting the
readings. Once the unit is warmed up, set the digital reading to 00.0 by pressing the
AUTO ZERO button. This may take several tries until the unit settles down. Avoid
any vibration or movement of the test head, since the unit is VERY sensitive. If the
7. Rem
Us
the drain tube from its storage clip. Holdin
tube just in front of the stopper with one hand while pulling the stopper out wit
other. Remove the stopper and allow the water to drain comp
to fully drain the unit when testing is co
and other potential water related damage.
Dispose of any water left in the opened bottle of degassed and distilled water. The
water will absorb oxygen over time, render
Remove the tube from its clip and raise
the tube end. Remove drain cap using a
combination of pinching the tube just
d let the water flow out.
below the cap and a twisting motion.
Once the cap is removed keep the tube
pinched until the drain tube is in the
container. Move container below the unit
until an