UPM 2000
Netech Corporation
110 Toledo St, Farmingdale, NY 11735
http://www.Netech.org
30
UPM
Check by laying the hanger flat on a table.
e
as shown below.
2.
If a transformer is to be used, plug it into the jack located at the
lower left rear of the unit. Then plug the transformer into a wall
socket.
2000 Systems Check
1. Level the unit using the two adjustable feet in the front of the base of the unit.
The unit is level when the bubble is centered. Inspect the transducer cone for
1-800-547-6547
dents and clean if needed
. Do
not use a dented cone. Incorrect readings may
result.
Inspect the cone hanger to verify that the hanger is not bent or distorted.
Make minor adjustments as needed to
ensure that the hanger lays flat. Insert the cone hanger into the protruding post
ol s. Carefully install the cone on the transducer mount pin on the cone hanger
h
Note:
There are transformers available for all plug and voltage
configurations, which can be ordered separately.
3. Allow sufficient time for the UPM 2000 to fully adjust to room temperature
. If
the unit is not at room temperature, the transducer will be slowly warming or
cooling and will cause the readings to shift.
4.
Obtain a bottle of the distilled and degassed water prepared following the
procedures discussed earlier in this chapter,
Preparing the Distilled and Degassed
Water
.
Be sure that the water is at room temperature before continuing with set
up.
Cone Top
Cone
Mounting Pin
Transformer Unit Power Jack
Cone Mounting
Hole