UPM 2000
Netech Corporation
110 Toledo St, Farmingdale, NY 11735
http://www.Netech.org
28
1-800-547-6547
Operation
When operated according to the instructions given in this chapter, the UPM 2000
digital ultrasound power meter will quickly and accurately measure the energy
output of ultrasound d i
Operating Environment
The UPM 2000
must be allowed to adjust to room temperature
before the unit can be
operated. The relative humidity in the room should not exceed 90%.
o
Chapter 4
for additional information on safety, storage and shipping
ements for the UPM 2000.
When using the UPM 2000 to test ultrasound devices, follow the safety
y the ultrasound manufacturer.
•
Do not use any coupling medium other than degassed and distilled water
.
•
Do not operate the UPM 2000 for more than 1 hour during testing. Change
water after 45 minutes of use.
ve and dry the cone when the test is complete.
ev ces.
Refer t
requir
Precautions
precautions outlined b
Use of improper media can lead to reading variations of up to 10%.
•
Do not spill water on the front or rear panel.
•
Do not apply more than 30 W of input power to the UPM 2000.
•
Remo
•
Drain the unit completely after each test session.