Recommended Procedures and Precautions
When using the UPM 2000 to test ultrasound devices, follow the safety precautions
specified by the ultrasound manufacturer.
Do not use any coupling medium other than degassed and distilled
water.
Do not spill water on the front or rear panel.
Do not apply more than 30 W of input power to the UPM 2000.
Do not operate the UPM 2000 for more than 1 hour during testing.
Change the water between extended tests.
Calibration Check Procedures
UPM 2000 calibration can be verified using the 100 gram weight supplied with the
unit to insure that the unit is in good working order.
This procedure is NOT
intended to substitute for a regular calibration.
The weight supplied is an OMIL
class M1 weight that will weigh between 99.7 and 100.3 grams.
1.
The unit should be off at the beginning of this procedure. Remove
the cone. Place the calibration weight adaptor pedestal on the
cone-mounting pin. This provides stable platform upon
which the 100 gram check weight will be placed.
2.
Turn the unit on and allow it to warm up for 5 minutes.
3.
Press the MODE button until the unit displays in grams.
4.
Zero the display.
5. Place the calibration check weight on the pedestal. Allow the
reading to settle down for 30 seconds to 1 minute. Take the
reading.
100
Gram Weight
Calibration Weight
Adapter Pedestal