MultiPro 2000 User Manual
33
Specifications
Specifications are subject to change without notice
Leakage Current:
All Current measurement loads are user selectable
to AAM ES1 or IEC60601-1 Load. The meter readings
correspond to the RMS value of the current
measured.
Measurement Mode: True RMS
Test Load: AAMI ES1 or IEC 60601
Range: Auto range: 0 to 99.9, 100 to 2000 uA.
Accuracy:
DC to 100 Hz: + 1% of the r 2 uA.
100 Hz to 1 kHz: + 2% of the r 3 uA.
1 kHz to 1 MHz: + 10% of the r 10uA.
Current Range:
0 – 99.9, 100 - 2000 Micro-Amps.
Power Cord Ground Resistance:
Measurement Mode: Two Wire
Test Current: 100 mA DC.
Range: 0 to 99 and 100 to 2400 MilliOhms.
Accuracy:
+ 2% of the r 2 MilliOhms.
Mains Line Voltage Measurement:
Range: 90 Volt to 255 V rms.
Accuracy: +
2% of r2 V
Device Current:
Range: 0.1 A( RMS) AC to 20 A C RMS.
Accuracy: + 2% of r0.1 A