MultiPro 2000 User Manual
17
Measurements
Earth Leakage:
Earth leakage is compliant to IEC 60601 and ANS/AAMI
ES1 standard. The default compliant standard is set to
AAMI ES1. The Compliant standard can be changed by
pressing F1 to toggle between (AAMI “ES1” or “IEC”
60601). This will be saved to the memory until it is
changed again.
There are four Leakage measurement options.
•
Normal Polarity
•
Reverse Polarity
•
Open Neutral
•
OFF. Select the desired polarity.
The Earth leakage measurement is through the internal
ground conductor and the external test lead is not
required. To measure the current flowing in the
protective Earth circuit, press the key marked “Case
Leakage”. The display will indicate “CASE LEAKAGE” and
“Select Polarity” and “ES1” /IEC (AAMI ES1) or “IEC”
60601) selection.
Select the desire polarity and press “READ”. The display
will indicate the leakage as well as the polarity selected
as shown in Screen Shot 3. Press F1 (“SAVE”) to save the
results.