MultiPro 2000 User Manual
19
Case Leakage: Using External Leads
Case /Enclosure leakage measurement is similar to
Earth Leakage measurement except the leakage
current is measured through the external test Lead.
To measure the case leakage/Enclosure leakage,
attach the external Kelvin cable to the case
(conductive surface) of the (DUT) device under test.
Select the desired polarity and press “READ”.
The display will indicate the leakage as shown in the
Screen Shot 4. Press F1 (“SAVE”) to save the results.
Case leakage measurements can be made for
various line conditions, power on, off, normal or
reverse polarity, closed or open ground, Closed or
open neutral.