Leads to Ground leakage:
Press F1 “GND” and select the desired leads from the
next menu. Press “READ” and the leakage current will be
displayed. Press F1 (“SAVE”)to save the results.
Press “ESC” to go back to the previous menu.
The block diagram on the following page represents the
electrical connection.
Leads Isolation leakage:
Caution:
In this test an isolated AC voltage equal to or 10% higher
is applied to the leads and earth ground. The voltage is
current limited but it is lethal. DO not touch the exposed
metal parts of the ECG post or the analyzers reference
ground.
To measure the isolation leakage current, select “ISO”
(F3) and press “READ”. The leakage current will be
displayed as shown.
CAUTION
:
In Isolation leakage test, the Analyzer
injects, current limited isolated AC voltage between the
earth ground and the patient leads of the DUT (patient
monitoring device). Although the voltage is current
limited it is lethal.