Section 7: Disassembly Guide
7-7
7.6
REPLACING THE FLOW SYSTEM
1.
Open the monitor case as described in section 7.2 "Opening the
Monitor Case".
The Flow System has 5 lines, refer to Figure 7-7: Flow System.
Two lines go from the pump:
•
Main line
•
Exhaust line
Three lines go from the solenoid:
•
Zero line
•
Input line 1
•
Input line 2 (not viewed in Figure 7-7)
Figure 7-7: Flow System