Section 7: Disassembly Guide
7-4
7.3
REPLACING THE CO
2
BOARD
Refer to Figure 7-4: Replacing CO
1.
Open the monitor case as described in section 7.2 "Opening the
Monitor Case".
2.
Remove the two CO
2
board Mounting Screws.
3.
Remove the CO
2
Connector Screw from the CO
2
Input
Connector.
4.
Remove the ON-OFF Button and save for reassembly.
5.
Pull out and remove the Gas Outlet and save for reassembly.
6.
Remove the CO
2
board and replace with a new CO
2
board and
reassemble in reverse order.
Note: When replacing the keypads, be sure the icons are facing the correct
direction. (Refer to Figure 1-1: Handheld Capnograph/Pulse
Oximeter.)
Figure 7-4: Replacing CO
2
board and Housing Components
Summary of Contents for NPB-75
Page 3: ...Handheld Capnograph Pulse Oximeter Service Manual 005554B...
Page 4: ...This page intentionally left blank...
Page 11: ...This page intentionally left blank...
Page 13: ...This page intentionally left blank...
Page 29: ...This page intentionally left blank...
Page 33: ...This page intentionally left blank...
Page 49: ...This page intentionally left blank...
Page 51: ...This page intentionally left blank...