Section 7: Disassembly Guide
7-3
2.
Place the monitor on a clean, dry, nonabrasive surface with the
back of the monitor facing up.
3.
Remove the two Bottom Mounting and two Back Mounting
Screws from the rear panel of the monitor as shown in
Figure 7-2: Removing Mounting Screws.
.
Figure 7-2: Removing Mounting Screws
4.
Carefully lift the rear case cover up and remove the 14-pin flat
cable from the CO
2
board as shown in Figure 7-3: Separating
Figure 7-3: Separating Rear and Front Covers