Section 5: Troubleshooting
5-4
Symptom
Probable Cause
Corrective Action
LED display
does not light
when the unit is
functioning.
The Power
Management option is
set at
LOW
.
Refer to the monitor’s
operator’s manual to
change the Power
Management option
to
NORMAL
or
HIGH
.
An
LED display
has
failed.
Replace with new
LED display.
A CO
2
board component
has failed.
Replace with a new
CO
2
board.
Beeper does not
beep.
The Pulse Tone option
on the Alarm Silence
menu has been
disabled.
Turn the Pulse Tone
ON. Refer to the
monitor’s operator’s
manual.
One or more of the
Alarm Silence options
has been turned
OFF
.
Turn the respective
alarm
ON
. Refer to
the monitor’s
operator’s manual.
A CO
2
component has
failed.
Replace with a new
CO
2
board.
Table 5-2: Advisory Messages Guide
Message
Probable Cause
Corrective Action
Check Unit
!
A CO
2
board component
has failed.
Check Event
Messages screen in
Service Mode.
Check CO
2
!
Check SpO
2
!
An SpO
2
board
component has failed.
Replace with a new
SpO
2
board.
Table 5-1: Troubleshooting Guide (Continued)
Summary of Contents for NPB-75
Page 3: ...Handheld Capnograph Pulse Oximeter Service Manual 005554B...
Page 4: ...This page intentionally left blank...
Page 11: ...This page intentionally left blank...
Page 13: ...This page intentionally left blank...
Page 29: ...This page intentionally left blank...
Page 33: ...This page intentionally left blank...
Page 49: ...This page intentionally left blank...
Page 51: ...This page intentionally left blank...