Section 6: Flow Calibration Check
6-2
6.3
FLOW CALIBRATION PROCESS
To perform the Flow Calibration Process, refer to Table 6-1: "Flow
Calibration".
Table 6-1: Flow Calibration
Objective
Action
Response
To access Flow
Calibration screen
After accessing
Service Level 2
(after entering the
password)
Simultaneously
long press
and
To change the
parameter setting or
value
short press
To move to the next
parameter (and save
the changed setting)
short press
To select Flow Cal.
short press
To start Calibration
short press
short press
Wait until either a Pass or Fail message is displayed.