Section 5: Troubleshooting
5-5
Table 5-3: Event Messages
Event Message
Probable Cause
Corrective Action
CO
2
board
A CO
2
board component
failed.
Replace with a new
CO
2
board.
SpO
2
board
Replace with a new
SpO
2
board.
Flow
- Loose connections
- Kinks or twists in
tubing
- Flow system blocked
- Inspect all tubing
ends for loose
connections and
ensure tight
connections
- Straighten kinks
and undo any twists
in the tubing.
- Check flow rate and
if necessary change
Flow System.
- Pump Malfunction
- Replace pump
- A CO
2
board
component has failed
- Replace with a new
CO
2
board