SECTION 2: REQUIRED EQUIPMENT
2.1
Required Equipment
2.1
REQUIRED EQUIPMENT
You will need the following equipment to disassemble, replace parts,
check, adjust, or calibrate the monitor.
For disassembly and replacement:
• Phillips-head screwdriver (medium)
• Flat-head screwdriver (small)
• IC Extractor for PLCC socket
• Threadlocker glue e.g. Loctite
®
222 super screw lock
For checks, adjustments and calibration:
• Calibration Gas (5% CO
2
in air)
• FilterLine or FilterLine H
• DS-100A - Durasensor
®
, Nellcor
• Leak Test Jig
• Vacuum manometer with a range of 0-500 mBar
e.g. MPB-3200-DM, Digital Manometer MicroPneumatic Logic, Inc.
• Communication Adapter Kit (only necessary to verify print functions)
• Seiko DPU-414 printer (only necessary to verify print functions)
• External power supply
• Handheld Capnograph/Pulse Oximeter Operator’s Manual
• Flow Meter (range 0-50 ml/min) e.g. Top Track Mass Flow Meter,
model: 822-13-OV1-PV1-V1, Sierra Instruments, Inc.
Warning: Observe ESD (electrostatic discharge) precautions
when handling, adjusting or performing any procedure with
the monitor’s internal components.