4.2.4 Setting Exposure Mode
Sets the exposure mode.
Format 2 CMD
□
VAL1 CR
CMD inm
VAL1
0,1,2
<Example>
inm
□
0 CR (Setting the exposure mode free run)
>OK
>inm 0
4.2.5 Setting Exposure Time
There are three types (int/inta/intb) on setting exposure time as follows.
Int:
【
Sets the exposure time for both line A and line B
】
int
Format 3 CMD
□
VAL1
□
VAL2 CR
CMD
int
VAL1 1(fixed)
VAL2 *914
~
1048575(Setting counter value)
Note:*The ranges of these counter values vary
according to output data formats as follows.
(85MHz at the output data rate setting)
[8192pixels_8tap] [4096pixels_4tap]
914
~
1048575
[8192pixels_4tap] [4096pixels_2tap] 1874
~
1048575
[8192pixels_2tap]
3922
~
1048575
(40MHz at the output data rate setting)
[8192pixels_8tap] [4096pixels_4tap] 2010
~
1048575
[8192pixels_4tap] [4096pixels_2tap] 4187
~
1048575
[8192pixels_2tap]
8538
~
1048575
<Example>
int
□
1
□
8500 <CR>
(
Setting exposure time 100
μs
)
>OK
>int 1,8500
int
:
Setting the exposure time for both line A and line B
inta
:
Setting the exposure time of line A
intb
:
Setting the exposure time of line B