User & Service Manual
XLTEK Trex HD Amplifier
47
Model Number
Description
8000SL-3
Sensor, Reusable, Soft, Large, 1 m (3 ft) cable
7000A
Flexi-Form® III Adult, 1 m (3 ft) cable, 24-pack 7000P
7000I
Flexi-Form III Infant, 1 m (3 ft) cable, 24-pack
7000N
Flexi-Form III Neonate, 1 m (3 ft) cable, 24-pack
6000CA
Sensor, Disposable, Adult, 45 cm (17.5 in) cable
6000CP
Sensor, Disposable, Pediatric, 45 cm (17.5 in) cable
6000CI
Sensor, Disposable, Infant, 90 cm (35.5 in) cable
6000CN
Sensor, Disposable, Neonate, 90 cm (35.5 in) cable
6500SA
Sensor, Durafoam Disposable, Standard, 1 m (3 ft) cable
6500MA
Sensor, Durafoam Disposable, Small, 1 m (3 ft) cable
11.6. Measurements
This pulse oximeter module is designed to determine the percentage of arterial oxygen saturation of
functional hemoglobin.
Pulse rate and SpO2 values are updated every 1/3 second and consist of a 4-beat average.
Measurement Wavelengths and Output Power
3
Red: 660 nanometers @ 0.8 mW max. avg.
Infrared: 910 nanometers @ 1.2 mW max. avg. (using Nonin PureLight® sensor)
Factors that may degrade pulse oximeter performance or affect the accuracy of
the measurement include the following:
excessive ambient light
excessive motion
electrosurgical interference
blood flow restrictors (arterial catheters, blood pressure cuffs, infusion lines, etc.)
moisture in the sensor
improperly applied sensor
incorrect sensor type
poor pulse quality
venous pulsations
3