NI DIDS-2100 Series User Manual | © National Instruments | 39
6. When the restoration process is completed, press
Exit
button to close the
Real-Time System Backup
dialog.
7. After the image restoration process, the DI Driver System will be rebooted automatically. Wait for the
USER1
and
USER FPGA
LEDs to start blinking, indicating the DI Driver System application is running.
8. If calibration files were backed up in step 1, use an FTP client to restore any .cdl calibration files. The .cdl
calibration files on the system controller should be located at “C:\ni-rt\LabVIEW Data\”. After restoring the calibration
file, reset the controller or cycle the power.