If you do not want to perform an adjustment, you can update the calibration date and onboard
calibration temperature without making any adjustments:
1.

Call the DAQmx Initialize External Calibration function to start a calibration session on
the FieldDAQ device. The default password is 

NI

.

2.

Call the DAQmx Set Temperature FieldDAQ Calibration function to input the external
temperature in degrees Celsius.

3.

Call the DAQmx Close External Calibration function to end the session. Set the Action
input to Commit.

Reverification

Repeat 

Verification

 to determine the As-Left status of the device.

Note

 

If any test fails reverification after performing an adjustment, verify that you

have met the all conditions listed in 

Test Conditions

 before returning your device to

NI. Refer to 

Where to Go for Support

 for assistance in returning the device to NI.

Accuracy Under Calibration Conditions

The values in the following table are based on calibrated scaling coefficients, which are stored
in the onboard EEPROM.

The following accuracy table is valid for calibration under the following conditions:

Ambient temperature 23 °C ± 5 °C

No nodes share power with the FieldDAQ device under calibration

The test limits listed in Tables 3 and 6 are derived using the values in the following table.

Table 8. FD-11613/11614 Voltage Accuracy Under Calibration Conditions

Mode

±PPM of Reading

±PPM of Range

*

 

 

 

High Resolution

362

55.9

Best 50 Hz Rejection

365

56.9

Best 60 Hz Rejection

365

56.9

High Speed

487

62.3

* Range = 78.125 mV

Note

 

Values in this table are intended for calibration verification only. These

values only apply under the specific calibration conditions described in this
document, and are not to be construed as general operational specifications of the
FD-11613 or FD-11614. For operational specifications, refer to the most recent

FD-11613 Specifications

 or 

FD-11614 Specifications

 at 

ni.com/manuals

.

FD-11613/11614 Calibration Procedure

  | 

© National Instruments

  | 

9

Summary of Contents for FD-11613

Page 1: ...to Go for Support 10 Software Requirements Calibrating the FD 11613 or FD 11614 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI DAQmx driver Earliest...

Page 2: ...pplications that use the NI DAQmx driver NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties Test Equipment The following table lists the equipment requi...

Page 3: ...rt Note You must install NI DAQmx 18 1 or later for device calibration support 2 Set up the FieldDAQ device as outlined in the FD 11613 11614 Quick Start 3 Configure the FieldDAQ device in Measurement...

Page 4: ...of operation and test points required to verify the FieldDAQ device The verification procedures assume that adequate traceable uncertainties are available for the calibration references Complete the...

Page 5: ...Accuracy Under Calibration Conditions 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage channel on the FieldDAQ device according to the follow...

Page 6: ...om each channel and average the readings e Clear the task 13 Compare the average to the limits in the following table Table 6 FD 11613 11614 Voltage Verification Test Limits for Zero Test Points ADC T...

Page 7: ...eldDAQ device as shown in the following figure Figure 3 Bank 1 Voltage Channel Adjustment Connections TC0 TC0 HI LO TC1 TC1 TC6 TC6 TC7 TC7 Calibrator FD 11613 11614 TC2 TC2 TC3 TC3 TC4 TC4 TC5 TC5 Ba...

Page 8: ...gh g for each reference value in the array of adjustment points 4 Disconnect the calibrator from the FieldDAQ device 5 FD 11614 Connect the calibrator to Bank 2 of the FD 11614 as shown in the followi...

Page 9: ...onditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the followin...

Page 10: ...d Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents coverin...

Reviews: