background image

d.

Read the samples and average the readings.

e.

Clear the task.

6.

Set the calibrator to Standby mode (STBY).

7.

Compare the average to the limits in Table 3.

8.

Repeat steps 3 through 7 for each test point.

9.

Repeat steps 3 through 8 for each ADC timing mode on the FieldDAQ device (High
Resolution, Best 50 Hz Rejection, Best 60 Hz Rejection, and High Speed).

10. Disconnect the calibrator from the FieldDAQ device.
11. Short all TC+ and TC- terminals on the FieldDAQ device together.
12. Acquire and average samples.

a.

Create and configure an AI voltage channel on the FieldDAQ device according to
Table 3.

b.

Configure the AI voltage channel timing according to Table 4.

c.

Start the task.

d.

Read the samples from each channel and average the readings.

e.

Clear the task.

13. Compare the average to the limits in the following table.

Table 6. FD-11613/11614 Voltage Verification Test Limits for Zero Test Points

ADC Timing

Mode

Range (mV)

Test Point

1-Year Limits (mV)

Minimum

Maximum

Location

Value

(mV)

Lower

Limit

Upper

Limit

 

 

 

 

 

 

 

High
Resolution

-78.125

78.125

Zero

0

-0.0044

0.0044

Best 50 Hz
Rejection

-78.125

78.125

Zero

0

-0.0045

0.0045

Best 60 Hz
Rejection

-78.125

78.125

Zero

0

-0.0045

0.0045

High Speed

-78.125

78.125

Zero

0

-0.0049

0.0049

The test limits in this table are derived using the values listed in 

Accuracy Under Calibration

Conditions

.

14. Repeat steps 12 through 13 for each ADC timing mode on the FieldDAQ device.
15. Disconnect the short from the TC channels.
16.

(FD-11614)

 Connect the calibrator to Bank 2 of the FD-11614, as shown in the following

figure.

6

  | 

ni.com

  | 

FD-11613/11614 Calibration Procedure

Summary of Contents for FD-11613

Page 1: ...to Go for Support 10 Software Requirements Calibrating the FD 11613 or FD 11614 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI DAQmx driver Earliest...

Page 2: ...pplications that use the NI DAQmx driver NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties Test Equipment The following table lists the equipment requi...

Page 3: ...rt Note You must install NI DAQmx 18 1 or later for device calibration support 2 Set up the FieldDAQ device as outlined in the FD 11613 11614 Quick Start 3 Configure the FieldDAQ device in Measurement...

Page 4: ...of operation and test points required to verify the FieldDAQ device The verification procedures assume that adequate traceable uncertainties are available for the calibration references Complete the...

Page 5: ...Accuracy Under Calibration Conditions 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage channel on the FieldDAQ device according to the follow...

Page 6: ...om each channel and average the readings e Clear the task 13 Compare the average to the limits in the following table Table 6 FD 11613 11614 Voltage Verification Test Limits for Zero Test Points ADC T...

Page 7: ...eldDAQ device as shown in the following figure Figure 3 Bank 1 Voltage Channel Adjustment Connections TC0 TC0 HI LO TC1 TC1 TC6 TC6 TC7 TC7 Calibrator FD 11613 11614 TC2 TC2 TC3 TC3 TC4 TC4 TC5 TC5 Ba...

Page 8: ...gh g for each reference value in the array of adjustment points 4 Disconnect the calibrator from the FieldDAQ device 5 FD 11614 Connect the calibrator to Bank 2 of the FD 11614 as shown in the followi...

Page 9: ...onditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the followin...

Page 10: ...d Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents coverin...

Reviews: