background image

Keep connections to the device as short as possible. Long cables and wires act as
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the device are secure.

Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire
to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 ±5 °C. The device temperature will be greater
than the ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the FieldDAQ device
measurement circuitry is at a stable operating temperature.

Calibration Procedure

The calibration process includes the following steps:
1.

Initial Setup

 on page 3

2.

Verification

 on page 4

3.

Adjustment

 on page 7

4.

EEPROM Update

 on page 8

5.

Reverification

 on page 9

Initial Setup

Complete the following steps to set up the FieldDAQ device.
1.

Install the software and NI-DAQmx driver as outlined in the 

FD-11613/11614 Quick

Start

.

Note

 

You must install NI-DAQmx 18.1 or later for device calibration support.

2.

Set up the FieldDAQ device as outlined in the 

FD-11613/11614 Quick Start

.

3.

Configure the FieldDAQ device in Measurement & Automation Explorer (NI MAX) as
outlined in the 

FD-11613/11614 Quick Start

.

4.

If the FieldDAQ device is not reserved automatically, select the device and click the

Reserve Network Device

 button. Refer to 

Reserving the Device in MAX

 for more

information.

5.

Self-test your device in MAX by expanding 

Devices and Interfaces

 »

Network Devices

 ,

right-clicking your FieldDAQ device, and selecting 

Self-Test

. Self-test performs a brief

test to determine successful device installation. When the self-test finishes, a message
indicates successful verification or if an error occurred. If an error occurs, refer to 

ni.com/

support/daqmx

.

Reserving the Device in MAX

When the FieldDAQ device is connected to a network, multiple users can access the device.
To perform any DAQ functionality on the device, including reset and self-test, you must
reserve the device in MAX. In MAX, an unreserved device or device reserved by another host

FD-11613/11614 Calibration Procedure

  | 

© National Instruments

  | 

3

Summary of Contents for FD-11613

Page 1: ...to Go for Support 10 Software Requirements Calibrating the FD 11613 or FD 11614 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI DAQmx driver Earliest...

Page 2: ...pplications that use the NI DAQmx driver NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties Test Equipment The following table lists the equipment requi...

Page 3: ...rt Note You must install NI DAQmx 18 1 or later for device calibration support 2 Set up the FieldDAQ device as outlined in the FD 11613 11614 Quick Start 3 Configure the FieldDAQ device in Measurement...

Page 4: ...of operation and test points required to verify the FieldDAQ device The verification procedures assume that adequate traceable uncertainties are available for the calibration references Complete the...

Page 5: ...Accuracy Under Calibration Conditions 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage channel on the FieldDAQ device according to the follow...

Page 6: ...om each channel and average the readings e Clear the task 13 Compare the average to the limits in the following table Table 6 FD 11613 11614 Voltage Verification Test Limits for Zero Test Points ADC T...

Page 7: ...eldDAQ device as shown in the following figure Figure 3 Bank 1 Voltage Channel Adjustment Connections TC0 TC0 HI LO TC1 TC1 TC6 TC6 TC7 TC7 Calibrator FD 11613 11614 TC2 TC2 TC3 TC3 TC4 TC4 TC5 TC5 Ba...

Page 8: ...gh g for each reference value in the array of adjustment points 4 Disconnect the calibrator from the FieldDAQ device 5 FD 11614 Connect the calibrator to Bank 2 of the FD 11614 as shown in the followi...

Page 9: ...onditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the followin...

Page 10: ...d Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents coverin...

Reviews: