Table 3. FD-11613/11614 Voltage Verification Test Limits for Positive and Negative Test

Points

ADC Timing

Mode

Range (mV)

Test Point

1-Year Limits (mV)

Minimum Maximum

Location

Value

(mV)

Lower

Limit

Upper

Limit

 

 

 

 

 

 

 

High
Resolution

-78.125

78.125

Negative FS

-70

-70.030

-69.970

Positive FS

70

69.970

70.030

Best 50 Hz
Rejection

-78.125

78.125

Negative FS

-70

-70.030

-69.970

Positive FS

70

69.970

70.030

Best 60 Hz
Rejection

-78.125

78.125

Negative FS

-70

-70.030

-69.970

Positive FS

70

69.970

70.030

High Speed

-78.125

78.125

Negative FS

-70

-70.039

-69.961

Positive FS

70

69.961

70.039

The test limits in this table are derived using the values listed in 

Accuracy Under Calibration

Conditions

.

4.

Set the calibrator to Operate mode (OPR).

5.

Acquire and average samples.
a.

Create and configure an AI voltage channel on the FieldDAQ device according to
the following table.

Table 4. FD-11613/11614 Voltage Channel Configuration

Physical Channel

Input Range (mV)

Units Terminal Configuration

Minimum Maximum

 

 

 

 

 

FD11613-Bank1/ai0:7 or
FD11614-Bank1/ai0:7

-78.125

78.125

Volts

Differential

b.

Configure the AI voltage channel timing according to the following table.

Table 5. FD-11613/11614 Voltage Channel Timing Configuration

ADC Timing Mode

Sample Mode

Samples to Read

Rate (S/s)

Timeout (s)

 

 

 

 

 

High Resolution

Finite

20

1.8

30

Best 50 Hz Rejection

Finite

80

7.1

30

Best 60 Hz Rejection

Finite

100

8.3

30

High Speed

Finite

1,000

85

30

c.

Start the task.

FD-11613/11614 Calibration Procedure

  | 

© National Instruments

  | 

5

Summary of Contents for FD-11613

Page 1: ...to Go for Support 10 Software Requirements Calibrating the FD 11613 or FD 11614 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI DAQmx driver Earliest...

Page 2: ...pplications that use the NI DAQmx driver NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties Test Equipment The following table lists the equipment requi...

Page 3: ...rt Note You must install NI DAQmx 18 1 or later for device calibration support 2 Set up the FieldDAQ device as outlined in the FD 11613 11614 Quick Start 3 Configure the FieldDAQ device in Measurement...

Page 4: ...of operation and test points required to verify the FieldDAQ device The verification procedures assume that adequate traceable uncertainties are available for the calibration references Complete the...

Page 5: ...Accuracy Under Calibration Conditions 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage channel on the FieldDAQ device according to the follow...

Page 6: ...om each channel and average the readings e Clear the task 13 Compare the average to the limits in the following table Table 6 FD 11613 11614 Voltage Verification Test Limits for Zero Test Points ADC T...

Page 7: ...eldDAQ device as shown in the following figure Figure 3 Bank 1 Voltage Channel Adjustment Connections TC0 TC0 HI LO TC1 TC1 TC6 TC6 TC7 TC7 Calibrator FD 11613 11614 TC2 TC2 TC3 TC3 TC4 TC4 TC5 TC5 Ba...

Page 8: ...gh g for each reference value in the array of adjustment points 4 Disconnect the calibrator from the FieldDAQ device 5 FD 11614 Connect the calibrator to Bank 2 of the FD 11614 as shown in the followi...

Page 9: ...onditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the followin...

Page 10: ...d Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents coverin...

Reviews: