nCounter Pro Analysis System User Manual
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
81
Continue following the instructions in each step of the workflow. Using lint-free paper, thoroughly
wipe off any excess silicone grease from the nozzles and tip ejectors. Be careful not to get grease
into
any nozzles. If grease does get in the nozzle, open up a paperclip and use the wire to carefully
scrape the grease back out.
4.
At the end of the workflow (Figure 123), press
finish to return to the Maintenance Menu.
Completing the Lubricate O-rings process will
reset the internal timer and remove the yellow
alert icon if it is present.
Figure 123: The ‘Lubricate O-rings—STEP 5’
screen.