SEV-00184-01
nCounter Pro Analysis System User Manual
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
37
Reporter Library Files
Users need only upload Reporter Library Files (RLFs) when they first receive a new CodeSet. If the
Enterprise Package is configured, RLFs may also be transferred automatically by FTPS (certain
configurations may disable transfer by USB flash drive). The RLF corresponding to a specific
CodeSet must be uploaded before setting up a run using that CodeSet. Once an RLF has been
uploaded, it will always be available until it is manually deleted from the Digital Analyzer.
Instructions for deleting RLFs are in
Technical Support and Maintenance
Upload RLFs via USB Flash Drive
1.
Insert the USB flash drive into the USB port
on the right side of the display.
2.
From the Main Menu, press upload files
(Figure 50).
3.
The ‘Upload Files’ screen will appear
(Figure 51). Select upload RLF.
Figure 50: The Main Menu.
Figure 51: The ‘Upload Files’ screen