nCounter Pro Analysis System User Manual
SEV-00184-01
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
54
4.
The Digital Analyzer will return to the
‘Cartridge information’ screen (Figure 72).
Press fov count to change the data resolution.
The ‘Select Data Resolution’ screen will appear
(Figure 73).
5.
Select the desired data resolution. Press enter.
The number of images (fields of view, or FOVs) collected corresponds to the sensitivity and
dynamic range that can be achieved. If a higher throughput is preferred on the analyzer, the user
may opt to collect fewer images if sensitivity is less important. For example, rescanning a
cartridge to check for data quality at low resolution takes approximately 2.5 minutes per flow cell
(lane)
vs.
20 minutes per flow cell at maximum resolution.
Resolution
Images per sample
low
25
medium
100
high
280
very high
490
max
555
NOTE:
Specifications for this system are based on the “high” data resolution setting.
Therefore, the “very high” and “max” resolution settings do not guarantee that the
throughput specification will be met.
Figure 72: The ‘Cartridge information’ screen
Figure 73: The ‘Select Data Resolution’ screen