nCounter Pro Analysis System User Manual
SEV-00184-01
FOR RESEARCH USE ONLY. Not for use in diagnostic procedures.
78
4.
Close the door and the electrode fixture will tilt
back to its home position. Press next.
Two screens will appear in succession, ending
on Step 5 (Figure 118).
5.
Press finish to return to the Maintenance
Menu.
Lubricate O-Rings
CAUTION:
When conducting maintenance activities, always use appropriate safety
precautions including wearing safety glasses and gloves.
ADVERTISSEMENT:
Pendant les tâches de maintenance, toujours mettre en œuvre les
mesures de précaution appropriées, notamment le port de gants et de lunettes de
sécurité.
The O-rings on the pipetting nozzles in the Prep Station are present to ensure a good seal with the
pipette tips, thus ensuring accurate liquid volume control. O-rings are the small black rings found
on the lower end of the nozzle. These O-rings must be lubricated once a month to maintain a good
seal.
The Prep Station Maintenance Menu has an interface to make O-ring lubrication easier. Following
the user interface brings the pipetting head to the front of the deck for access during lubrication.
Required materials:
•
Silicone grease (transparent grease in a plastic container, provided by NanoString)
•
Lint-free paper (e.g., Kimwipe)
•
Gloves
•
Pipette tip
Figure 118: The ‘Align Electrodes—STEP 5’ screen.