53
Calibration
5.5
Calibration Types: Zero vs. Span
The TG5000 has two types of calibration: Zero and Span calibration.
Zero Calibration resets the baseline level reading to zero. If the target gas is suspected to be occa-
sionally present, it is best to also use a zero gas cylinder during the zero calibration. If the target
gas is not present in the atmosphere, an additional calibration cylinder is not required.
The "Calibrate" option involves first applying a zero gas followed by the span gas. The span gas
is a known concentration of gas which adjusts the accuracy and precision of the transmitter to the
known value; this is referred to as the “Span Value. See Fig. 25.
Fig. 25
Calibration Curve
The Sensor Span Value in the device menu should be the same as the concentration listed on the
calibration gas cylinder; unless an LEL simulant gas is being used.
The XIR Plus sensor can be calibrated to a wide variety of gas compounds using either 0.1 %
propane, 0.6 % propane, or 2.5 % methane, and MSA's gas table. See Tab. 21 for a complete list
of gas compounds and corresponding tables and span values.
US
Summary of Contents for TG5000
Page 1: ...Operating Manual TG5000 Gas Monitor Order No 10212126 L Y REV 0 US ...
Page 36: ...36 Operation US ...
Page 46: ...46 Operation US ...
Page 62: ...62 Maintenance US ...
Page 63: ...63 Maintenance US ...
Page 64: ...64 Maintenance US ...
Page 70: ...70 Appendix Calibration Guide and Additional Gases US ...
Page 71: ...71 Appendix Calibration Guide and Additional Gases R 32 US ...
Page 72: ...72 Appendix Calibration Guide and Additional Gases US ...
Page 75: ...For local MSA contacts please visit us at MSAsafety com ...