Identify the sensor assembly needed via the A-5K-SENS code on the interior sensor label and
obtain the appropriate sensor assembly. Screw the replacement XCell sensor onto the Sensor
Body Assembly, ensuring that the XCell sensor mates flush against the bottom edge of the Sensor
Body Assembly.
Fig. 26
Removing the XCell Sensor
Fig. 27
Digital Sensor - Position of Label
Alarm set points, span value, full scale limit and alarm direction will not change when
replacing a sensor with the same gas and range. Alarm set points, span value, full scale
limit and alarm direction will change to the new sensor settings when replacing a sensor
with a different gas and/or range.
The TG5000 Gas Monitor is shipped with the Sensor Swap Delay enabled. This means that the
4-20 mA output signal and the fault relay will hold off a fault indication for two minutes before the
missing sensor indication is displayed on the device. This setting allows the operator to
exchange sensor modules without a fault indication. See Swap Delay in section 4.2.1 for more
details.
It is recommended that all other maintenance be performed at an MSA
factory-authorized service center.
US
ATO code label
Summary of Contents for TG5000
Page 1: ...Operating Manual TG5000 Gas Monitor Order No 10212126 L Y REV 0 US ...
Page 36: ...36 Operation US ...
Page 46: ...46 Operation US ...
Page 62: ...62 Maintenance US ...
Page 63: ...63 Maintenance US ...
Page 64: ...64 Maintenance US ...
Page 70: ...70 Appendix Calibration Guide and Additional Gases US ...
Page 71: ...71 Appendix Calibration Guide and Additional Gases R 32 US ...
Page 72: ...72 Appendix Calibration Guide and Additional Gases US ...
Page 75: ...For local MSA contacts please visit us at MSAsafety com ...