Care and maintenance
30
Install the sterile filter with caution to avoid damage to the filter or the unit.
-
Remove the old sterile filter by turning the sterile filter counterclockwise from the
designated port
-
To avoid any leaks, wrap the threads of the sterile filter with teflon tape
-
Attach the new sterile filter at the dispenser by turning the sterile filter clockwise in
the dedicated port
-
At first water dispensing please open the deaeration screw for a short time to vent
the sterile filter (see )
Modul replacement
7.3
Figure 13: Modules in opened system
-
Remove the left side panel
-
For exchanging of ProPak-R10 or UPW-Pak follow the steps listed in chapter 5.2.1
-
For the removal of the tubes from the modules and the reattachment of the new
-
After the flushing and resetting the operating hours counter, the system
automatically returns to the normal operation state
-
At first water dispensing please open the deaeration screw for a short time to vent
the sterile filter (see )
-
Check impermeability on all connections
-
After mounting the side panel, the system is fully operational again
The highest ultra pure water quality is only reachable, if the resin is swollen. This
Reverse osmosis
housing
ProPak-R10
(black
arrows)
UV reactor
UPW-Pak
(blue
arrows)