Errors and Countermeasures
28
Trouble
Reason
Measure
hours.
(18) OPERATING HOUR
LIMIT ProPak. Exchange
necessary.
The ProPak-R10
(Pretreatment module)
exceeded the maximum
operating hours.
Replace the ProPak-R10
(pretreatment module).
Reset the expired operating
hours.
(25) No Pressure, check
water input!
No feed water pressure.
Check the main water
supply. Check the pressure
sensor. Inform the customer
service.
(30) Q3 is out of range!
Check Q3 sensor.
30 mins Q3 is less then 0.04
µS/cm
Inform the customer service.
(31) Tank level is incorrect!
Please check tank level
sensor!
The tank level sensor (top )
is ON, but
(bottom) is OFF. This could
only possible if the bottom
sensor is defect.
Inform the customer service.
Table 3: Display warning messages