Care and maintenance
34
Consumables
7.7
Description
Article no.
Sterile filter
190-0013
ProPak-R10
290-0065
UPW-Pak (
organic
applications)
UPW-Pak (
anorganic
applications)
190-0087
190-0088
UV lamp
921-0138
Tank vent filter
450-0025
Table 5: Consumables
System update
7.8
In order to update the system with newer software, it requires a USB-Stick which is not
bigger than 4 GB and should be formatted in
.ext
-Format. The new software should be
available in the USB-Stick which formatted in
.mip
file.
-
Restart system with an available update in USB-Stick
will open this window
-
Click “OK” to continue
-
Software-version will be displayed:
o
HW-Version: actual in the system
o
Flash Ver.: newer version in USB-Stick
-
Click “Yes” to continue
-
Click “No” will cancel the update request
-
System will update automatically. Please do not turn
off the system.