SECTION 1
6
Specifications
Feature
Specifications
Instrument Type
12-lead ECG and SpO
2
(T12S only) digital transmitter
Input Channels
Continuous 12-lead signal acquisition with SpO
2
acquisition (T12S
only) and transmission
ECG Leads Transmitted
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5 and V6
Pulse Oximetry Transmitted
(T12S only)
Plethysmographic waveform, oxygen saturation percent (SpO
2
%),
pulse rate and probe status spot check during hookup; continuous
transmission during monitoring
SpO
2
% Accuracy**
70% - 100% SpO
2
: E
RMS
= + 2.2%
SpO
2
Pulse Rate Accuracy**
45 – 210 BPM: E
RMS
= + 2%
SpO
2
Measurement wavelength
Red:
660nm at 3.5 to 4.5 mW nominal
Infrared:
905nm at 3.5 to 4.5 mW nominal
Frequency Range*
608.48 MHz to 631.52 MHz or
904.76 MHz to 925.16 MHz or
2400.96 MHz to 2482.56 MHz
Special Functions
Lead impedance check, ECG display, and SpO
2
check (T12S only)
during hookup, lead fail, probe fault and battery notification, multi-
purpose call; 10-wire, 5-wire, and 4-wire options
Water Resistance
Electronics protected from water damage; meets EN 60529
rating IPX2
Defibrillator Protection
Complies with AAMI standards and IEC 60601-2-25
Number of Channels
256 user selectable
Function Keys
Up, down, and enter keys for on/off and menu navigation; dedicated
call button during transmission
Device Classification
Type BF (SpO
2
)/CF (ECG), battery operated
Weight
4.6 oz. (130 g) without battery
Dimensions
3” x 4.4” x .98” (76 mm x 112 mm x 25 mm)
Battery
1 AA alkaline, 24-hour typical life
* Operating frequency range is dependent on the T12/T12S part number.
** Verified by clinical tests where measured values of the sensors were compared with arterial
CO-oximetry in adult subjects over the specified SpO
2
range. Because pulse oximeter equipment
measurements are statistically distributed, only about two-thirds of pulse oximeter equipment
measurements can be expected to fall within the
E
RMS
of the value measured by a CO-oximeter.
NOTE: Manufacturer does not supply accessories for direct cardiac applications.
NOTE: Operating range performance is 15 meters with no barriers.
NOTE: Disposable sensors are recommended for use on patients weighing more than 66 lbs./30 kg.
NOTE: Reusable wrap sensors are recommended for use on patients weighing more than 22 lbs./10 kg.
NOTE: Reusable clip sensors are recommended for use on patients weighing more than 44 lbs./20 kg.