SECTION 2
9
Turning the T12/T12S On
The T12/T12S will power up as soon as a battery with a minimum of 1.0 volts has been inserted into the battery
compartment and the cover applied. If the T12/T12S was turned off after its last use, the user has two options to
power the T12/T12S on:
1.
Remove and reapply the top cover, or
2.
Press the
Up
button.
The T12/T12S will power up and display the initial startup menu within three seconds.
Turning the T12/T12S Off
The user has two options to power the T12/T12S off:
1.
Remove the top cover, or
2.
Press and hold the
Up
button for a period of three seconds
a. A prompt will appear in the LCD display
b. Press
the
Down
button to highlight Yes
c.
Press the
Call
button to select Yes
NOTE: The user has five seconds to select yes or no or the prompt will be replaced by the main menu and
the power-off operation is cancelled.
Patient ECG Hookup
Skin Preparation
Skin preparation is important to perform before electrode attachment to help ensure good signal quality when
transmitting patient data. Poor skin-electrode contact may cause noise or artifact which can affect the analysis of the
ECG data. Low amplitude signals may also be the result of poor skin-electrode contact.
To prepare the skin
1.
Identify the electrode sites on the torso by referring to
Positioning the Electrodes
.
2.
Remove any hair from the electrode sites using a razor.
3.
Wipe oils from the electrode sites with an alcohol prep pad.
4.
Remove any dead skin from the electrode sites with an abrasive pad. Two to three moderate rubs at each site
should be sufficient.