Monark 928 G3 / 928 G3r
17
Astrand
(Except 928 G3r)
The test is automatically stopped after six minutes if
Steady State (SS) is active. If not, the test continues
until SS is reached or the time reaches 12 minutes.
Steady State (SS) is checked by comparing the average
HR during the period 4:45 to 5:00 with the average
HR during the period 5:45 to 6:00. If the difference
between these two average values is the same or less
considered SS.
When the test is completed the results are calculated
and shown directly in the display.
You can stop the test manually after six minutes by
pressing START STOP even if SS is not active (an
error message is shown). The result is calculated and
displayed.
References / Literature:
Astrand P-O, ”Ergometri - konditionsprov”, Monark, Sverige
•
Åstrand I, ”Aerobic work capacity in men and women with special
•
reference to age”, Acta Physiol Scand. 49 (suppl. 169), 1960
Astrand P-O, ”Experimental studies of physical working capacity in
•
relation to sex and age”, Munksgaard, Köpenhamn, 1952.
Astrand P-O, Rodahl K, ”Textbook of Work Physiology”, McGraw-
•
Hill, New York, 1970.
Explanations to error messages:
LO Hr
HI Hr
Err
No SS
The end pulse is too low and outside
the Astrand tables so no results can be
calculated.
The end pulse is too high and outside
the Astrand tables so no results can be
calculated.
Another reason why no results can be
obtained.
The test is cancelled manually without
SS after 6 minutes or no SS at time 12
minutes.