68
0112-0151 K
To adjust dispense accuracy for a program
Dispense accuracy in the AquaMax Microplate Washer is set to
specification during manufacturing. However, various installation
configurations can also affect accuracy, such as location and size of
reagent bottles, types of reagent, and so on. These adjustment are
specific and individual to every program. However, after you determine
an adjustment for a given set of parameters, you can use the same
setting whenever that combination is used (Liquid type, Head type, and
so on).
The AquaMax Microplate Washer supports 4 L or 10 L bottles only.
For best results test dispense accuracy in a separate program and use the
results to adjust your working program.
1.
Make sure that the system is set up to match the parameters of
your working program.
2.
Create a new program and in the
Program Settings
dialog, touch
Program Options
.
3.
Touch
Adjust Dispense
.
Figure 4-8:
Adjust dispense options
4.
In the
Adjust Dispense
dialog, set the adjustment to 0% for the
chosen fluid inlet.
5.
Weigh an empty plate.
6.
Complete the program with a dispense function from the fluid
inlet of choice: 300
μ
L/well/96-well plate or 100
μ
L/well/384-well
plate.
7.
Run the program.
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...